Mapleson Classification

In general anaesthesia, recovery and emergency resuscitation procedures involve the use of a breathing system. The principles of anaesthetic systems is to efficiently eliminate exhaled carbon dioxide, without greatly increasing dead space or resistance. In 1954 Mapleson classified breathing systems based on their efficiency in eliminating carbon dioxide during spontaneous respiration. The breathing systems are classified in order of increased requirement of fresh gas flow (FGF) to prevent rebreathing during spontaneous respiration. System A requires 0.8 - 1 times, B and C require 1.5 - 2 times and D, E and F require 2 - 4 times of the patient's minute volume.
a This system consists of a reservoir bag close to the fresh gas outlet with the APL valve placed at the patient end (Magill A, or expiratory limb - Parallel Lack). During inspiration, as the patient's inspiratory flow is greater than the fresh Gas Flow, additional gases will be drawn out from the reservoir bag. During the expiratory phase, exhaled gases will force the fresh gas back into the circuit and reservoir bag. This will result in increasing the pressure in the breathing system and thus forcing the exhaled gases to escape through the APL valve (scavenging port). the system requires a low flow (0.8 - 1 times of Minute Volume, approximately 6-8 litres/minute) during spontaneous respiration to prevent rebreathing. This system is not recommended for intermittent positive pressure ventilation (IPPV).
b In this system the reservoir bag, fresh gas supply and APL valve are closer to the patient. This will cause mixing of inspiratory and expiratory gases and therefore a higher flow rate (1.5 - 2 times of the patient's minute volume, i.e. 12 - 16 litres/min) is required to prevent rebreathing during spontaneous respiration. Due to the risk of rebreathing and reduced delivery of oxygen rich gases to the patient this system is no longer used.
b This system is similar to Mapleson B, however, the bag is positioned very close to the patient. This system is used for manual ventilation during resuscitation. A flow rate of 1.5 - 2 times of the minute volume is required to avoid rebreathing.
b This system is mainly used with spontaneous respiration. The Fresh Gas Flow is close to the patient and the APL valve is placed away from the patient. The risk of rebreathing in this circuit will be high especially in patients who have a short expiratory pause or do not have an expiratory pause (infants). To overcome this problem a high FGF of 2 - 4 times of the patient's Minute Volume is required to prevent all rebreathing. In practice some rebreathing is tolerated and in an adult an FGF of 6 - 7 litres/minute will maintain a normal arterial CO2 tension
b This system is primarily for use in neonates and paediatrics, where low resistance is of great importance. There is no APL valve (to reduce resistance) and a high FGF, 2 - 4 times of the patient's Minute Volume (with a minimum flow of 3 litres/minute) is required to eliminate rebreathing risk during spontaneous ventilation. This system is recommended for up to 22kg (approximate tidal volume of 140ml).
b Jackson Rees made a great improvement to T-Pieces (Mapleson E) by adding an open tail 0.5 litre reservoir bag to the end of the expiratory (reservoir) limb. This allows manual ventilation and the application of Positve End Expiratory Pressure (PEEP) to help maintain open airways. For controlled ventilation, normocapnia can be maintained with a FGF of 1000ml + 100ml/kg body weight