Airway Management

Anaesthesia Breathing Systems

Anaesthesia Components

Critical Care

Neonatal Breathing Systems

Breathing Filters

Oxygen & Aerosol Therapy

Resuscitation

Suction

Urology & Continence Care

Colostomy

Enema Tips

Mapleson Circuits

 

In general anaesthesia, recovery and emergency resuscitation procedures involve the use of a breathing system. The principle of anaesthetic systems is to efficiently eliminate exhaled carbon dioxide, without greatly increasing dead space or resistance. In 1954 Mapleson classified breathing systems based on their efficiency in eliminating carbon dioxide during spontaneous respiration. The breathing systems are classified in order of fresh gas flow (FGF) to prevent rebreathing during spontaneous respiration. Further information on the Mapleson classification can be found in our educational section.

 

Mapleson A - (Magill, Parallel Lack and Coaxial Lack)
This system consists of a reservoir bag close to the fresh gas outlet with the APL valve placed at the patient end (Magill A, or expiratory limb-Parallel Lack). During inspiration, as the patient's inspiratory flow is greater than the Fresh Gas Flow, additional gasses will be drawn out from the reservoir bag. During the expiratory phase, exhaled gases will force the fresh gas back into the circuit and reservoir bag. In the expiratory phase, exhaled gases will force the fresh breathing system and thus forcing the exhaled gasses to escape through the APL valve (scavenging port). The system requires a low flow (0.8-1 times of Minute Volume, approximately 6-8 litres/ minute) during spontaneous respiration to prevent Rebreathing. This system is not recommended for intermittent positive pressure ventilation. (IPPV)

 

Mapleson C - (Waters Bag, Bagging System - Adult, Direct and Paediatric and Resuscitation Bag).
This system is similar to Mapleson B, however, the bag is positioned very close to the patient. This system is used for manual ventilation during resuscitation. A flow rate of 1.5-2 times of the minute volume is required to avoid Rebreathing.

 

Mapleson D - Coaxial Bain
This system is mainly used with ventilated patients. The Fresh Gas Flow is close to the patient and the APL valve is placed away from the patient. The risk of Rebreathing in this circuit is high especially in patients who have a short expiratory pause or do not have an expiratory pause (infants). To overcome this problem a high FGF of 2 - 4 times of the patient's Minute Volume is required to prevent all Rebreathing. In practice some rebreathing is tolerated and in an adult FGF of 6 - 7 litres/minute will maintain a normal arterial CO2 tension.

 

Argentic Systems
Argentic is a range of anti-microbial breathing systems to combat microbial contamination within the clinical setting and reducing the risk of Hospital Acquired Infections (HAI). Argentic breathing systems are impregnated with a silver ion additive which rapidly reduces the number of bacterium. When oxygen in the ambient air comes into contact with silver particles it creates a catalytic reaction which helps to inactivate the microbes.

argentic-circuit.gif

 

 

Product CodeDescriptionPack SizeImage
038-01-256 Mapleson A Magill System (1.6m)
22mm lightweight corrugated tube with 2L Reservoir Bag and Adult APL Valve
20
038-01-256Ag As above with Argentic Tubing 20
038-01-257 Mapleson A Parallel Lack System (1.6m)
22mm lightweight corrugated tubes with 2L Reservoir Bag and Adult APL Valve
20
038-01-257Ag As above with Argentic Tubing 20
038-01-255 Mapleson C Bagging System (1.6m)
10mm lightweight corrugated tube with 2L Reservoir Bag and Adult APL Valve + 15F/6-9mm adaptor
20
038-01-258 Mapleson C Bagging System (2.1m)
Star Lumen tube with 2L Reservoir Bag and Adult APL Valve + 22mm Female adapter
20
038-01-251 Mapleson D Bain System (1.6m)
22mm lightweight corrugated tube with 2L Reservoir Bag and Adult APL Valve
20
038-01-251Ag As above with Argentic Tubing 20
038-01-253 Mapleson D Bain System (2.4m)
22mm lightweight corrugated tube with 2L Reservoir Bag and Adult APL Valve
20
038-01-253Ag As above with Argentic Tubing 20
038-01-254 Mapleson D Bain System (5.4m)
22mm lightweight corrugated tube with 2L Reservoir Bag and Adult APL Valve
20
038-01-254Ag As above with Argentic Tubing 10
038-01-200 Basic Bain Coaxial System (1.6m)
22mm lightweight, flexible corrugated tube with 22mm M swivel connector + low profile elbow
20
038-01-200Ag As above with Argentic Tubing 20
038-01-210 Basic Bain Coaxial System (2.4m)
22mm lightweight, flexible corrugated tube with 22mm M swivel connector + low profile elbow
20
038-01-210Ag As above with Argentic Tubing 20
038-01-201 Basic Bain Coaxial System (1.6m)
22mm lightweight, low compliance corrugated tube with 22mm M swivel connector + low profile elbow without serrated connector
20
038-01-201Ag As above with Argentic Tubing 20
038-01-200B Basic Bain Coaxial System (1.6m) with limb
22mm lightweight, flexible corrugated tube with 22mm M / 15mm F swivel connector + 2L Reservoir Bag + low profile elbow + two spare M/M connectors + 1.6m limb
20
038-01-200BAg As above with Argentic Tubing 20