LuerSafe Breathing Filter |

LuerSafe breathing filters incorporate
a unique cap-less self sealing luer port which will only accept the
correct corresponding colour coded LuerSafe monitoring line. This
in turn will open the valve and provide uninterrupted flow to
sample CO2 levels when connected.
LuerSafe eliminates erroneous misconnections to standard luer
connections such as IV tubing or other connectors ensuring a
correct connection to the monitoring port only.
|
Cap-less Design
The self sealing valve prevents air flow in and out of the filter
when a monitoring line is not connected, eliminating the need to
have a cap to be manually fixed and removed.
Colour Coded Valve
The LuerSafe Valve is visibly seen through the filter housing by
its distinctive orange colour which can be easily matched to the
correct corresponding colour coded monitoring line connector.
Unique Design Luer
Port
Eliminating the risk of misconnection by not accepting a standard
luer connector. The unique size only fits the LuerSafe monitoring
line identified by the colour coded connector.
|
 |
|

To connect a CO2 monitoring line to a LuerSafe filter,
locate the orange colour coded LuerSafe Monitoring
Line Connector.
|

Insert the connector into the LuerSafe Port and rotate
in a clockwise direction, depressing the valve until a secure
lock is reached.
|
|

Verify that CO2 readings are visible
|

To remove, unscrew the monitoring line and the valve
will self seal.
|